• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS Back to Search Results
Model Number FD-411UR
Medical Device Problem Code Output above Specifications (1432)
Health Effect - Clinical Code Fever (1858)
Date of Event 01/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is related to the following linked patient identifiers: (b)(6) and (b)(6).The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that during a therapeutic endoscopic submucosal dissection under sedation, using the electrosurgical forceps, the blend cut current "burned" despite lowering all the parameters and changing the effects.As a result, the physician completed the dissection procedure using power coagulation with the same device.Post-procedure the patient developed thermocoagulation syndrome, presenting with fever, which required the administration of intravenous fluid and antibiotics.The patient's symptoms are currently in remission, and their condition is stable.
 
Event or Problem Description
No additional information received from customer.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The complaint was not confirmed due to no device return.The cause cannot be established due to no findings available.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS
Common Device Name
SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21413296
Report Number9614641-2025-00162
Device Sequence Number8317419
Product Code KNS
UDI-Device Identifier04953170251542
UDI-Public04953170251542
Combination Product (Y/N)N
PMA/510(K) Number
K062517
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFD-411UR
Device Lot NumberB001761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer01/31/2025
Initial Report FDA Received Date02/19/2025
Supplement Report FDA Received Date02/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ELECTROSURGICAL GENERATOR; ELECTROSURGICAL KNIFE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-