Catalog Number 4251644-02
Medical Device Problem Code Fail-Safe Problem (2936)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was returned for evaluation.The defect cannot be confirmed due to not having any evidence to support the reported incident.Retained units were evaluated and passed the internal testing.Since a sample was not returned for evaluation, the exact root cause of this incident could not be determined.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or nonconformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Event or Problem Description
As reported by uf/importer report # (b)(4): describe the event or problem: emergency department (ed) registered nurse (rn) was placing peripheral intravenous access line (piv) in patient.After successful piv access was obtained, rn removed needle and attempted to deploy needle safety mechanism.Safety only deployed 1/3 of the way up the needle.No injury reported to patient or rn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key21413988
Report Number2523676-2025-00154
Device Sequence Number21226543
Product Code FOZ
UDI-Device Identifier04046963166007
UDI-Public(01)04046963166007(11)240708(17)290630(10)0061928058
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4251644-02
Device Lot Number0061928058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 01/28/2025
Initial Report FDA Received Date02/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexUnknown
-
-