EXACTECH, INC. ADVANCED PATELLA 35MM 3 PEG IMPLANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
200-07-35 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
1038671-2023-00853, 1038671-2025-01097.D10: (b)(6) - 02-022-45-5040 - truliant tib fit tray cem sz 5f / 4t, (b)(6) - 02-022-35-5009 - truliant tib imp ps insert sz 5 9mm, (b)(6) - 02-020-11-0350 - truliant ps cem fem ps cem right sz 5.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00853, 1038671-2025-01097.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 52 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, osteolysis, pain, scar tissue, swelling/edema, and synovitis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: g3/g4, h4, h6.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|