• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US CLRCANNULASMOOTH/DISTALRIB7X75MM5/PK -ST; CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US CLRCANNULASMOOTH/DISTALRIB7X75MM5/PK -ST; CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 214115
Medical Device Problem Codes Delivered as Unsterile Product (1421); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: investigation summary :the complaint device was discarded by the customer, therefore not returned to j&j medtech orthopaedics and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (2311133), and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
This is report 5 of 7 for (b)(4).It was reported that prior to a shoulder arthroscopy procedure, a total of seven cannula devices were found to be completely damaged.The reporter noted that the boxes were bent and compromised, and that the sterility of certain cannulas had been affected due to tears.While the outer implant boxes were damaged, the packaging inside remained sterile.The reporter stated that they chose to retain the implants since they were undamaged.The cannulas were discarded after being taken out of trunk stock.There was a slight delay to the surgical procedure.However, the duration of the delay was not specified.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLRCANNULASMOOTH/DISTALRIB7X75MM5/PK -ST
Common Device Name
CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key21414594
Report Number1221934-2025-00628
Device Sequence Number8523444
Product Code GEA
UDI-Device Identifier10886705003563
UDI-Public(01)10886705003563(10)2311133(17)281130
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214115
Device Lot Number2311133
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2025
Supplement Date Received by Manufacturer03/07/2025
Initial Report FDA Received Date02/19/2025
Supplement Report FDA Received Date03/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-