| Catalog Number |
9-ASD-017 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 17mm amplatzer septal occluder was selected for implant.Using an unknown sized abbott delivery system.During procedure, the occluder deformed with a cobra shape in the left atrial disc.There were no interactions with cardiac structures during deployment and no angulation/kink was observed with the delivery system.The occluder was never released from the delivery cable while inside the patient and removed.The device was exchanged for a new 17mm amplatzer septal occluder, which was successfully implanted.The patient was reported to be in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed via returned device analysis.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Additionally, a review of the complaint history identified no similar incidents reported from this lot.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|