| Model Number |
H7493941838300 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported via medwatch that stent dislodgement occurred.The target lesion was located in the ostial left circumflex coronary artery.The lesion was pre-dilated with a small nc balloon.A 3.00 x 38mm synergy xd drug-eluting stent was introduced for treatment but experienced advancing difficulty through the left main stent.The stent came off of the stent delivery system within the vessel.A balloon was then advanced to the distal end of the stent in attempt to remove them as one unit.The stent was removed.There were no patient complications reported as a result of this event.
|
| |
|