• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION, INC. LIFEBAND® LOAD-DISTRIBUTING BAND; CARDIAC RESUSCITATOR BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION, INC. LIFEBAND® LOAD-DISTRIBUTING BAND; CARDIAC RESUSCITATOR BAND Back to Search Results
Model Number 100
Medical Device Problem Codes Material Integrity Problem (2978); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll has not received the lifeband in complaint for investigation.A follow-up report will be filed if and when the product is returned and investigation has been completed.The death was not related to the autopulse device.The autopulse is used as an adjunct to manual cpr, adjunctive use-only indication is prominently displayed on device labels and in the device manual.The benefit of using autopulse is that it in part substitutes mechanical compressions for the physical labor of manual chest compressions when effective manual cpr is not possible.If the autopulse did not start or unexpectedly stops compressions, rescuer should revert to manual cpr, which is the standard of care.The autopulse is intended to be used as an adjunct to manual cpr on adult patients.In case of stoppage of autopulse the trained user reverts to manual cpr.The transition from autopulse to manual cpr by trained users is similar to the time necessary for rescuer rotation and presents the same workflow as manual cpr.Hence, based on available information, the patients' outcome was not negatively impacted by the interruptions when compared to standard-of-care manual cpr.According to multiple studies, out-of-hospital cardiac arrest (ohca) is considered one of the leading causes of death in industrialized nations, with a significant number of deaths occurring outside of a hospital setting, making it a major public health concern; this information is supported by research from the cdc, studies on global mortality rates, and data from various cardiac arrest registries.Survival to hospital discharge after non-traumatic emergency medical services-treated cardiac arrest with any first recorded rhythm was around 10% for patients of any age (mozaffarian, circulation, 2016; ebiomedicine 2023).Death is an expected outcome for ohca.
 
Event or Problem Description
The autopulse platform (sn (b)(6)) along with lifeband (lot # unknown) were used in an attempt to resuscitate a patient in cardiac arrest.The customer reported that the patient was properly aligned, and clothing was removed, but the lifeband repeatedly twisted.The customer stated no error codes were recorded, but the crew switched to manual cpr.The customer did not provide further details about the cardiac arrest event but reported that the patient passed away.The customer believes the issue was not due to user error and indicated that the patient's death was unrelated to the autopulse system.The customer mentioned that the reported issue was replicated at the station after the call but did not specify whether it was related to the autopulse platform or the lifeband.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEBAND® LOAD-DISTRIBUTING BAND
Common Device Name
CARDIAC RESUSCITATOR BAND
Manufacturer (Section D)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa sackrison
2000 ringwood ave.
san jose, CA 95131
4084192347
MDR Report Key21417564
Report Number3010617000-2025-00128
Device Sequence Number9244794
Product Code DRM
UDI-Device Identifier00849111001571
UDI-Public00849111001571
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100
Device Catalogue Number8700-0701-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2025
Initial Report FDA Received Date02/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-