| Catalog Number |
7757-2MP65/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The account alleges that the catheter broke at the hub during a procedure in the heart.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device has been returned for evaluation.The device was visually and microscopically investigated.The complaint is confirmed.The root cause is attributed to the manufacturing process.A review of the device history record was performed and no exception documents were found.A review of the complaint database was performed and one similar complaint for this lot from the same account was identified.
|
| |
|