The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the device revealed two reprocessing marks.Inspection of the elbow junction and tubing to cuff adapter, revealed no separation.The quick connect o-ring and overall, integrity of the connector appeared to be in-tact with no visual indication of damage.The tubing was clear with no visual kinks, bends, or distress.The entirety of the cuff appeared to be intact with no obvious signs of rips, tears, or fraying.Pilling was observed on the velcro sleeve.The received complaint device was connected to the device under testing (dut) tubing set and checked for hermeticity.The ball valve was then opened to allow for the complaint device to fill and to stabilize.The ptc was not able to stabilize and a hiss/leak was heard coming from the bladder.A small laceration was found.The device inspection and functional testing indicated that the complaint was confirmed for the reported failure mode.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: ptc damaged or broken during use (leak).Product damaged during shipping / storage.Inadequate handling instruction, user damages cuff.Loss of structural integrity.The instructions for use (ifu) state: warnings: it is important that the tourniquet cuff be applied at the proper location with adequate pressure for the appropriate amount of time.Avoid needles, towel clips, leg holders and other equipment that can puncture or otherwise damage the cuff.Directions for use: before beginning the procedure, verify compatibility of all devices and accessories.Prior to surgery, select the proper sized tourniquet cuff by measuring the circumference of the patient's limb.This will avoid problems caused by a tourniquet cuff that is too small or too large.Secure the cuff fasteners to ensure that the cuff stays in place during the procedure.If the package is damaged or if it was opened and the device was not used, return the device and packaging to stryker sustainability solutions.Inspect the device for overall condition and physical integrity.Do not use the device if any damage is noted.Return the device and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.The reported event will continue to be monitored through post-market surveillance.
|
|
It was reported that the tourniquet cuff would not inflate on any machine.As reported, the staff also tried different inflation tubes but it would not work.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|