| Model Number |
3286 |
| Medical Device Problem Codes |
High impedance (1291); Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Inadequate Pain Relief (2388)
|
| Date of Event |
01/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced ineffective therapy.Reprogramming has been unable to resolve the issue.Diagnostic testing revealed multiple impedance issues on both leads.As a result, surgical intervention may occurred on a later date.
|
| |
|
Event or Problem Description
|
|
Additional information indicates that patients leads had high impedances.Surgical intervention was undertaken on (b)(6) 2025 wherein leads were explanted and replaced to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction e1- initial reporter.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|