| Model Number |
5286 |
| Medical Device Problem Code |
Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an in clinic follow up, the device was interrogated, and invalid characters resulted in a device error.Technical support was contacted and a programmer interrogation would clear the data corruption.The patient was stable.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the device was restored to resolve this event.
|
| |
|
Search Alerts/Recalls
|