• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG 1688 AIM 4K AUTOCLAVABLE CAMERA HEAD C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG 1688 AIM 4K AUTOCLAVABLE CAMERA HEAD C-MOUNT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 1688410105
Medical Device Problem Code Fogging (1253)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was foggy image.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
Alleged failure: foggy image.Confirmed failure: deformed rear enclosure - replace rear enclosure.Damaged front enclosure threads - replace front enclosure.Intermittent cable - replace cable.Probable root cause: cables, hdmi cables, connectors, digital board, power supply, filter / fuse, ac inlet board, main board, transition board, intermittence between transition board and ch connector, software, camera head (sensor, sensor cable), coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp, zoom ring), problem toggling light source, connected monitors, leaking, poor grounding (e.G camera head connection from cable to transition board.), no/incorrect communication with light source, soaking cap disconnection during sterilization, electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge, cybersecurity attack, pendulum ch inertial measurement unit (imu), incident light from surgical scene is reflected by coupler/ch window, use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG 1688 AIM 4K AUTOCLAVABLE CAMERA HEAD C-MOUNT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key21426555
Report Number0002936485-2025-00129
Device Sequence Number7782018
Product Code GCJ
UDI-Device Identifier07613327562545
UDI-Public07613327562545
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K212511
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1688410105
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2025
Supplement Date Received by Manufacturer03/27/2025
Initial Report FDA Received Date02/20/2025
Supplement Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-