| Catalog Number |
ER8075 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the available information, during prep, all of the air did not come out of the reservoir and the physician suspected there was a leak.The reservoir was not implanted and a different 75 cc reservoir was placed instead.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this complaint, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Reservoir was received for evaluation.No functional abnormalities were noted with the reservoir.The information received indicated the air did not come out of the reservoir and a leak was suspected, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, during prep, all of the air did not come out of the reservoir and the physician suspected there was a leak.The reservoir was not implanted and a different 75cc reservoir was placed instead.
|
| |
|