|
DEPUY SYNTHES PRODUCTS LLC KINCISE¿ FEM HEAD IMPACTOR, REPLACE CAP; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
| Catalog Number |
1013-00-901 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.E1: the reporter¿s phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hip surgery, it was discovered that the tip of the fem head impactor device broke and cracked into two pieces.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11 additional narrative: h6: device history review: the actual device was not returned for evaluation; therefore, the reported condition was not confirmed.However, a device history review was performed which indicated that there were no non-conformances or abnormalities identified during the manufacture of the device that may have contributed to the reported condition.
|
| |
|
Search Alerts/Recalls
|
|
|