| Catalog Number |
362791 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® ppt¿ plasma preparation tubes k2e (edta) 9.0 mg that two (2) tubes exhibited floaties in the plasma (poor plasma).There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd had not received samples or photos for evaluation.Complaints for sample quality were not under statistical control for the month of january 2025, additional testing may be required: further clinical testing could not be conducted as the tubes were expired 30nov2024.Clinical evaluation cannot be conducted on expired tubes.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode poor plasma (floaties in plasma).Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® ppt¿ plasma preparation tubes k2e (edta) 9.0 mg that two (2) tubes exhibited floaties in the plasma (poor plasma).There was no health impact or consequence reported.
|
| |
|