C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT 8F CHRONOFLEX; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
1668362 |
| Medical Device Problem Code |
Expulsion (2933)
|
| Health Effect - Clinical Codes |
Erythema (1840); Necrosis (1971); Skin Erosion (2075)
|
| Date of Event |
04/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that sometimes post a port placement, the device allegedly had fifty percentage of the port eroded through the skin.It was further reported that the surrounding erythema was noted without induration, purulent drainage, or evidence of necrosis.However, the current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported port expulsion, skin erosion, erythema and necrosis as no objective evidence was provided for review.Furthermore, the clinical conditions alleged in the complaint cannot be confirmed.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: g3, h4, h6 (component, method).H11: d4 (unique identifier (udi) #), h6 (result, conclusion).H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that sometimes post a port placement, the device allegedly had fifty percentage of the port eroded through the skin.It was further reported that the surrounding erythema was noted without induration, purulent drainage, or evidence of necrosis.However, the current status of the patient was unknown.
|
| |
|