| Catalog Number |
J490G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 2/21/2025 h6 component code: g07002 - device not returned this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, the following was obtained: it was reported the following, - event occurred on (b)(6)2025 ((b)(4)) - were ".The sutures seemed to break off from the needle.We broke one suture on (b)(6)2025." - event occurred on (b)(6)2025 ((b)(4)) - were ".The sutures seemed to break off from the needle.He date we noticed was (b)(6)2025 we broke 6 that day." please confirm below for each event.- what was being done when the needle pulled-off (removal from package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? - is this device or samples available for product analysis? if yes, to whom should the shipper kit be sent? (please provide contact name, department, street address including zip, no po box please, email address, and phone number).- how many devices are available to be returned for product analysis? - provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep).For the first incident when the needle was pulled off we had the needle drivers on the needle, clamped and when the needle drivers were unclamped the suture detached from the needle itself.The other incident occurred when the technician was pulling through tissue, the needle went through but the suture did not.I do have 3 sutures left that can be analyzed for further information.The shipper kit can be sent to (please refer the attached email), email: (please refer the attached email), phone number: (please refer the attached email).I am also the external person to follow up with the sales rep.A device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that an animal underwent an animal procedure on an unknown date and suture was used.It was reported they had just opened a new box of suture for a dental appointment a week to two weeks ago.Unfortunately the sutures seemed to break off from the needle and then became unusable.They broke 7 out of the 12 sutures in this box.She only had 3 sutures left out of this box so they were only able to use 2 sutures successfully.The patient did not have any adverse reaction to the suture.The date we noticed was 1-14-25 we broke 6 that day and we broke one suture on (b)(6)2025.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 4/3/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Three unopened samples were received for analysis.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no pull off - suture needle was noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|