| Model Number |
ANGLED REAMER SLEEVE, 20° |
| Medical Device Problem Codes |
Mechanical Jam (2983); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by a sales representative via (b)(4) that an ar-9676 angled reamer drive shaft and an ar-9597-20 angled reamer sleeve fused together.This was discovered during an rtsa procedure and was completed with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Visual evaluation noted strong abrasion on the base of the angle reamer and damage inside the holes of the sleeve collar.Functional testing was not performed due to the returned part ar-9597-20 and ar-9676 were stuck and unable to separate.The most likely cause for the reported failure can be attributed to misuse due to a force perpendicular to the drive shaft being applied while the device is spinning.This applied force creates an excess of friction and heat that ultimately has the potential to cause the device to seize from functioning as intended.Application of this force perpendicular to the drive shaft is not needed for the device to function, nor is it recommended that the user do this during normal clinical use.When used normally, this issue is unlikely to present itself, which suggests that the handling of the device has a large impact on whether or not it will seize.For this reason, the direct cause is considered to be misuse of the device.(b)(4).Was opened to address this issue.
|
| |
|
Search Alerts/Recalls
|