DEPUY ORTHOPAEDICS INC US HOLDER FOR REAMING GUIDES; EXTREMITY INSTRUMENTS : HANDLES Back to Search Results
Catalog Number 230774001
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during intraoperative deflection of the proximal reamer guide size 16, there were difficulties in screwing the inner rod of the reamer guide into the inserter (to attach the reamer guide size 16 firmly to the instrument).It was not possible to unscrew it; the inner rod was bent and the tip was broken off (fragment did not fall into the site).No patient harm.Surgery time extended by several attempts to remove the reaming guide size 16, which was still in the humeral shaft, using the customer's own alternative instruments in order to insert the original implant.Surgery delay 20 minutes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary according to the information received, " during intraoperative deflection of the proximal reamer guide size 16, there were difficulties in screwing the inner rod of the reamer guide into the inserter (to attach the reamer guide size 16 firmly to the instrument).It was not possible to unscrew it; the inner rod was bent and the tip was broken off." the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation did not reveal that holder for reaming guides (230774001) had any disassembly issue.Review of provided photo shows the device however without having actual device for evaluation reported condition cannot be confirmed and cause remains undetermined.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the holder for reaming guides would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the device lot number is unknown, therefore a¿device history review could not be performed.¿ if the lot/serial number becomes available, the record will be re-assessed.Added: b5.
 
Event or Problem Description
Please confirm if there is any allegation against the 2 unk insertion devices as per sales rep the answer is "no".
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d4 (lot), d9, h4.Corrected: d4 (primary udi #).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, " during intraoperative deflection of the proximal reamer guide size 16, there were difficulties in screwing the inner rod of the reamer guide into the inserter (to attach the reamer guide size 16 firmly to the instrument).It was not possible to unscrew it; the inner rod was bent and the tip was broken off." the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation did not reveal that holder for reaming guides (230774001) had any disassembly issue.Review of provided photo shows the device however without having actual device for evaluation reported condition cannot be confirmed and cause remains undetermined.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the holder for reaming guides would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Device history batch: null.Device history review: null.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information provided: "instruments of (b)(4) were used again.During intraoperative deflection of the proximal reamer guide size 16, there were difficulties in screwing the inner rod of the reamer guide into the inserter (to attach the reamer guide size 16 firmly to the instrument).It was not possible to unscrew it; the inner rod was bent and the tip was broken off (fragment did not fall into the site).No patient harm.Surgery time extended by several attempts to remove the reaming guide size 16, which was still in the humeral shaft, using the customer's own alternative instruments in order to insert the original implant.Surgery delay 20 min".The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection revealed the overall holder for reaming guides surface with signs of heavy usage.No damage could be observed on its threaded surface nor in their subcomponents.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was performed for the holder for reaming guides using the returned rod for reaming guide holder as mating device.Functional test revealed the devices could not fully seat or assemble, as a consequence of the damage observed on the mating component's threads.Based on the condition observed, it is reasonable to confirm the reported allegation, as the damage observed on the mating device could have contributed to a difficulty or inability of both assembly and disassembly process.A manufacturing record evaluation was performed for the finished device 5244750 lot number, and no non-conformances nor manufacturing irregularities were identified.The overall complaint was confirmed as the observed condition of the holder for reaming guides would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: quantity manufactured: (b)(4).Date of manufacture: 27-may-2015.Any anomalies or deviations identified in dhr: no.Expiry date: n/a.Ifu reference: (b)(4).A manufacturing record evaluation was performed for the finished device 5244750 lot number, and no non-conformances nor manufacturing irregularities were identified.Device history review: a manufacturing record evaluation was performed for the finished device 5244750 lot number, and no non-conformances nor manufacturing irregularities were identified.H11 additional narrative: added: h4.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLDER FOR REAMING GUIDES
Common Device Name
EXTREMITY INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21432365
Report Number1818910-2025-02894
Device Sequence Number14543107
Product Code LXH
UDI-Device Identifier10603295116127
UDI-Public(01)10603295116127(10)5244750
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K170748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number230774001
Device Lot Number5244750
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2025
Supplement Date Received by Manufacturer02/25/2025
03/11/2025
03/28/2025
05/22/2025
Initial Report FDA Received Date02/21/2025
Supplement Report FDA Received Date02/27/2025
03/12/2025
03/31/2025
05/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-