| Model Number |
6481779 |
| Medical Device Problem Code |
Gas Output Problem (1266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The malfunction of compressor's drainage valve was discovered during the inspection of the ventilator.There was no patient harm.Manufacturer's ref.#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This event occurred on the indian market on a significantly similar device to ¿compressor mini 115v 60hz¿ which is sold in the us.Therefore, for d4 unique identifier (udi) #, primary di number has been used from compressor mini 115v 60hz.Provided in report: d4 serial # and h4 device manufacture date, correspond to device involved in the event, which is "compressor mini 230v¿.
|
| |
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by distributor's field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the drainage valve was found defective.The drainage valve is powered with 12 v and electrically controlled by the pc board.The purpose of the drainage valve is to remove water collected in the water separator at regular intervals.The valve is closed in non-activated condition.During operation, the valve will be activated (open) for approx.0.5 seconds twice every minute by the timing circuit on the pc board.When the valve opens, the water collected in the water separator will be transported to the drainage bottle.Water in the air gas module may damage the air gas module.Unfortunately, the claimed part has been not available for the further analyze of the issue.The cause of the issue was related to drainage valve.
|
| |
|
Event or Problem Description
|
|
Manufacturer's ref.#: (b)(4).
|
| |
|
Search Alerts/Recalls
|