| Catalog Number |
8065998330 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Visual Impairment (2138); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H.3., h.6.: investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that the surgeon questioned about the system readings and the intraocular lens was rotated by one hundred twenty degrees, as the result the patient had induced astigmatism of three diopters in the right eye of a patient, during cataract surgery.There are two related reports for this patient.This report addresses the patient's right eye and another manufacturer report will be filed for the fellow eye.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6., and h.11.A manufacturing device history record review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.Service history was reviewed for the system.There was no service record relevant to the reported event, found.All company surgical systems are verified to meet specifications after installation and customer-initiated servicing.Based on the information obtained, the root cause of the reported event is inconclusive.The manufacturer internal reference number is: (b)(4).
|
| |
|