|
Updated fields: d4, d9, g3, g6, h2, h3, h6, h11.The peritoneal catheter (id 823045) was returned for evaluation.Failure analysis - the catheter was visually inspected; no defects noted.The catheter was leak tested; a leak was observed.The catheter was irrigated, no occlusions noted.The complaint was confirmed.Root cause analysis - the issue reported by the customer could be confirmed.The possible root cause could be a sharp or pointed object coming into contact with the device or the stapler having damaged the catheter.As noted in the ifu, silicone has a low cut/tear resistance.
|