| Model Number |
1420 |
| Medical Device Problem Codes |
Connection Problem (2900); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the controller had poor/unstable mechanical connection due to bent pins in the power port.It is unknown if excessive force was used when trying to connect the controller.The controller was exchanged at home as the patient lives in a remote location and is on hospice care.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for investigation completion.Product event summary: the controller (b)(6) was not returned for evaluation.A review of the manufacturing documentation was not performed as the reported event is not related to a manufacturing or servicing issue.The reported event could not be confirmed due to insufficient evidence.Applicable risk documentation and experience with events of similar circumstances were considered; a possible root cause of the reported event can be attributed to misalignment during connection attempts between the metal receptacles on the controller and the plastic connector plugs in the battery and/or adapter cables.The socket pins may not withstand applied forces from subsequent misaligned connections, causing the pins to bend.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|