| Model Number |
APB-1.5-2-3D-ES |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id: apb-2-4-3d-es (229477270); implant date: n/a; explant date: n/a.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that two axium coils were prematurely detached.The patient was undergoing surgery for treatment of an amorphous, unruptured aneurysm in the c7 segment with a max diameter of 3.3 mm and a 1.2 mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was minimal.It was reported that during the aneurysm surgery, the coil was detached prematurely and did not completely embolized into aneurysm.After adjustment with a new coil and stent, the surgery was successfully completed.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that no detachment attempts were made.There was no kink/damage observed on the pushwire.The coil was implanted, but was not in the intended location.No additional medical/surgical intervention was required as it was pressed with a stent to stick it to the wall.Two coils had been implanted before and after the coil malfunctioned.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Product analysis: equipment used: video inspection system (m-85519), ruler (m-83361), camera (panasonic lumix dmc-zs5) as found condition: the axium prime device was returned for analysis within a shipping box; within a sealed plastic biohazard pouch; within an opened axium prime inner pouch; within a dispenser coil and within an introducer sheath.A second axium prime device was also returned and will be addressed in pli-20.Visual inspection/damage location details: the actuator interface was found securely attached to the coupler tube.The break indicator was found still intact.There were no evidence of detachment attempts at these locations.No damages or irregularities were found with the pusher.The coin was found still against the lumen stop.The shield coil was found undamaged.The implant coil was found still attached to the pusher.The implant coil was found damaged within the introducer sheath.Conclusion: based on the analysis performed, the customer report of ¿premature detachment" could not be confirmed as the implant was still attached.The implant coil was found damaged, indicative of potential advancing against resistance.H6.Coding updated based on analysis findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|