MEDTRONIC MEXICO S. DE R.L. DE CV AXIUM PRIME BARE PL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number APB-1.5-2-3D-ES
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: product id: apb-2-4-3d-es (229477270); implant date: n/a; explant date: n/a.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received a report that two axium coils were prematurely detached.The patient was undergoing surgery for treatment of an amorphous, unruptured aneurysm in the c7 segment with a max diameter of 3.3 mm and a 1.2 mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was minimal.It was reported that during the aneurysm surgery, the coil was detached prematurely and did not completely embolized into aneurysm.After adjustment with a new coil and stent, the surgery was successfully completed.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received reported that no detachment attempts were made.There was no kink/damage observed on the pushwire.The coil was implanted, but was not in the intended location.No additional medical/surgical intervention was required as it was pressed with a stent to stick it to the wall.Two coils had been implanted before and after the coil malfunctioned.
 
Additional Manufacturer Narrative
H3.Product analysis: equipment used: video inspection system (m-85519), ruler (m-83361), camera (panasonic lumix dmc-zs5) as found condition: the axium prime device was returned for analysis within a shipping box; within a sealed plastic biohazard pouch; within an opened axium prime inner pouch; within a dispenser coil and within an introducer sheath.A second axium prime device was also returned and will be addressed in pli-20.Visual inspection/damage location details: the actuator interface was found securely attached to the coupler tube.The break indicator was found still intact.There were no evidence of detachment attempts at these locations.No damages or irregularities were found with the pusher.The coin was found still against the lumen stop.The shield coil was found undamaged.The implant coil was found still attached to the pusher.The implant coil was found damaged within the introducer sheath.Conclusion: based on the analysis performed, the customer report of ¿premature detachment" could not be confirmed as the implant was still attached.The implant coil was found damaged, indicative of potential advancing against resistance.H6.Coding updated based on analysis findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BARE PL 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key21434361
Report Number9617601-2025-00201
Device Sequence Number4349779
Product Code KRD
UDI-Device Identifier00763000313647
UDI-Public00763000313647
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAPB-1.5-2-3D-ES
Device Catalogue NumberAPB-1.5-2-3D-ES
Device Lot Number229115240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2025
Supplement Date Received by Manufacturer02/24/2025
04/19/2025
Initial Report FDA Received Date02/21/2025
Supplement Report FDA Received Date03/18/2025
05/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
-
-