| Model Number |
35MM CENTRAL SCREW, MODULAR |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by a sales representative via phone that an ar-9560-24-2 arthrex univers revers modular glenoid system modulas baseplate and an ar-9561-35s univers revers modular glenoid system central screw fell apart once implanted.The surgeon used a screwdriver to remove the implants from the patient.The case was completed using a new baseplate and central screw.This was discovered during a revers shoulder procedure on (b)(6) 2025.The patient was not affected.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-9561-35s, 35mm central screw serial/batch number (b)(6), was received for investigation.A visual evaluation found damage and deformation in the internal hex connector hole.Functional testing was not conducted to avoid damage to the well-known mating part.Functional testing with the returned ar-9560-24-2 cannot be performed as the internal hex of the 35mm central screw was deformed.The most likely cause of the reported failure is user error, specifically, failure to achieve the appropriate torque requirements when tightening the central screw directions for use dfu-0189 univers revers¿ shoulder prosthesis system e.Warnings.8.Failure to achieve the appropriate torque requirements when tightening locking screws may result in the premature loosening of the device.22.Proper anchoring is of decisive importance for firm, permanent positioning of the prosthesis.
|
| |
|
Search Alerts/Recalls
|