MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE ICONIX 2.3MM REUSABLE DRILL; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE ICONIX 2.3MM REUSABLE DRILL; BIT, DRILL Back to Search Results
Catalog Number 3910-500-573
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the drill broke during the procedure.The broken pieces were retrieved.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
Updated e1 phone number to (b)(6).
 
Event or Problem Description
It was reported that the drill broke during the procedure.The broken pieces were retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: drill bit broken probable root cause: application -excessive force while drilling -selection of wrong drill -reuse of disposable drill -use of reusable drill past lifetime (after it has become dull) -starting and stopping drill -pulling drill out of bone without running drill -moving drill guide during drilling the reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the drill broke during the procedure.The broken pieces were retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICONIX 2.3MM REUSABLE DRILL
Common Device Name
BIT, DRILL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key21435449
Report Number0002936485-2025-00133
Device Sequence Number11913153
Product Code HTW
UDI-Device Identifier07613252598718
UDI-Public07613252598718
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-500-573
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by Manufacturer04/18/2025
04/18/2025
Initial Report FDA Received Date02/21/2025
Supplement Report FDA Received Date05/16/2025
05/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/18/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age43 YR
Patient SexMale
Patient Weight120 KG
-
-