| Model Number |
280025MM |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Fatigue (1849); Insufficient Information (4580)
|
| Date of Event |
01/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 2800 pericardial aortic valve underwent valve-in-valve intervention after an implant duration of nine (9) years, six (6) months due to unknown reasons.The tavr procedure was performed with a 26mm 9750tfx transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, b5, b7, d4, g3, g6, h2, h4, h6.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 2800 pericardial aortic valve underwent valve-in-valve intervention after an implant duration of nine (9) years, six (6) months due to aortic stenosis.The patient initially presented with fatigue and exercise intolerance.The tavr procedure was performed with a 26mm 9750tfx transcatheter valve.The patient was stable post procedure and then discharged home.
|
| |
|
Additional Manufacturer Narrative
|
|
Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.A definitive root cause is unable to be determined, however, patient and/or procedural factors likely caused or contributed.
|
| |
|