MAUDE Adverse Event Report: ARTHREX, INC. TIGHT CURVED SPEAR, FIBERTAK, REUSABLE; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TIGHT CURVED SPEAR, FIBERTAK, REUSABLE; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number TIGHT CURVED SPEAR, FIBERTAK, REUSABLE
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 01/27/2025, it was reported by a arthrex subsidiary employee via (b)(6) that an ar-3610ctc-2 fibnertak spear and an ar-3600 fibertak anchor became stuck in the drill guide when the surgeon went to insert.This occurred during use in an arcr case with no patient harm.This case was completed by using another product of same product number.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to misaligned insertion and/or prying/leveraging the device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIGHT CURVED SPEAR, FIBERTAK, REUSABLE
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21436540
Report Number1220246-2025-00599
Device Sequence Number19961430
Product Code LXH
UDI-Device Identifier00888867354296
UDI-Public(01)00888867354296(11)230918(10)0072329
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTIGHT CURVED SPEAR, FIBERTAK, REUSABLE
Device Catalogue NumberAR-3610CTC-2
Device Lot Number0072329
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2025
Supplement Date Received by Manufacturer09/04/2025
Initial Report FDA Received Date02/21/2025
Supplement Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-