It was reported that during a case on the (b)(6) 2025 when the bone harvesting kit for ria 2 was used, it was noticed that the plastic was coiling while reaming with the guidance of x-ray.The instrument was examined and it was noted that the particular yellow colored plastic attachment had burst inside the instrument hence causing it to not work properly.When tried to disassemble at the end of the operation, the damaged plastic had gotten stuck to the instrument and could not be removed.Another separately packed reaming kit was opened and used.
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6: the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation revealed that drive shaft f/ria 2 l520 (03.404.035) had broken.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces.The overall complaint was confirmed as the observed condition of the drive shaft f/ria 2 l520 would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to component failure and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : lot provided is not a valid number, therefore, the dhr could not be completed.If device is returned or lot number can be confirmed, the dhr will be revisited.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|