MAUDE Adverse Event Report: STRYKER GMBH ADVANCED LOCKING SCREW IMN SCREWS Ø5X55MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH ADVANCED LOCKING SCREW IMN SCREWS Ø5X55MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 23615055S
Medical Device Problem Codes Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Insufficient Information (4580)
Date of Event 10/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
As reported: "als is gradually loosening up after surgery".
 
Additional Manufacturer Narrative
Please note the corrections to d4 gtin, h6 clinical signs and h6 device code.The reported event could be confirmed based on the x-ray image[s] provided, only.A physical inspection was impossible since the item was not made available for inspection [implanted].Provided x-ray images were forwarded to an hcp for review and statement ¿ his comments - excerpts: the screw does back out, that is clear.Furthermore, images were forwarded to r&d for review and comments as well ¿ excerpts: possible causes: the thread has been damaged during implantation.The thread is worn due to postoperative stress.However, based on details available a more specific medical or technical statement could not be made and more detailed information about the complaint event as well as the affected device must be available in order to determine the exact root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise, we reserve the right to reopen the case.
 
Event or Problem Description
As reported: "als is gradually loosening up after surgery".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED LOCKING SCREW IMN SCREWS Ø5X55MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21439249
Report Number0008031020-2025-00354
Device Sequence Number4360428
Product Code HSB
UDI-Device Identifier07613327381610
UDI-Public(01)07613327381610(17)340331(10)K1AD5D3
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K193308
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number23615055S
Device Lot NumberK1AD5D3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2025
Supplement Date Received by Manufacturer04/03/2025
Initial Report FDA Received Date02/21/2025
Supplement Report FDA Received Date04/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-