| Catalog Number |
430840010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A physician reported that after direct selective laser trabeculoplasty procedure, the patient experienced unexpected increase in intraocular pressure (iop).Additional information was received as spike in iop and patient undergo with treatment medications.There are four medical device reports associated with this report.This is 3 of 4.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
The company representative was unable to confirm nor replicate anything that would have contributed to the reported event.The system was tested and found to meet product specifications.The customer reported event cannot be confirmed.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.Relevant complaints were found and reviewed as part of this investigation.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|