| Model Number |
3602536 |
| Medical Device Problem Codes |
Break (1069); Defective Component (2292); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding the devices used for spinal ther apy.It was reported that the female mechanism used to engage the t10 driver to lock the implant height is stripped.The tip of the driver is cracked.The middle set screw of crosslink as arrived in packaging was stuck, it would not loosen or tighten and upon trying to loosen the implant broke.There was no patient involvement reported.There were no further complications reported regarding the event.
|
| |
|
Search Alerts/Recalls
|