H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: one sealed 8.0 airguard peel-apart sheath and vessel dilator were received for evaluation.Visual evaluation was performed on the returned device.The sealed package was inspected thoroughly.Small holes were noted on the clear side of the sealed package.Components did not appear affected.No other anomalies were noted.The manufacturing site evaluation states, an initial view showed a sealed airguard kit, the ifu was observed inside the packaging, a tear was observed in the transparent part of the pouch near the supplier's seal, the components inside did not appear to be affected.A closer look at the additional images provided by field assurance showed two small holes in the packaging, the id of the holes were 0.4146mm and 0.7398mm, no other anomalies were observed throughout the sample.Therefore the investigation is confirmed for the reported tears, rips or hole in device packaging.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: g3, h4, h6 (method, component).H11: d4 (unique identifier (udi) #), h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|