• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TITAN SPINE ANTERALIGN¿ SPINAL SYSTEM WITH TITAN NANOLOCK¿; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TITAN SPINE ANTERALIGN¿ SPINAL SYSTEM WITH TITAN NANOLOCK¿; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 46221455
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6 - the device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from the health care professional via a manufacturer representative regarding a device used for spinal therapy.It was reported that the implant migrated.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Event or Problem Description
Additional information received from manufacturer representative that there is no plan/schedule for the revision surgery at this time.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Radiographic image review performed by dr hoit and the review comments are as per below 1.Ap fluoro lumbar interbody fusion anterior lateral interbody graft at all 3 levels present, posterior screw fixation 2.Anterior x-ray/fluoro the superior interbody graft appears interbody migrated in 1 versus 2 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTERALIGN¿ SPINAL SYSTEM WITH TITAN NANOLOCK¿
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
TITAN SPINE
6140 w. executive dr. suite a
mequon WI 53092
Manufacturer (Section G)
TITAN SPINE
6140 w. executive dr. suite a
mequon WI 53092
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21443239
Report Number3006340236-2025-00013
Device Sequence Number7939585
Product Code MAX
UDI-Device Identifier00763000128890
UDI-Public00763000128890
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number46221455
Device Catalogue Number46221455
Device Lot NumberTM0137628
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2025
Supplement Date Received by Manufacturer02/07/2025
03/24/2025
Initial Report FDA Received Date02/24/2025
Supplement Report FDA Received Date02/25/2025
03/28/2025
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
-
-