|
Updated fields - b4, g1, g3, g6, h1, h2, h3, h6 (type of investigation, investigation findings, component code, investigation conclusions), h11.A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse replaced the power supply.The fse replaced the blood detect tubing and clear 0.062" polyurethane tube because of a brown/yellow color.The fse replaced the helium tank valve knob because the valve clamping chain was broken.Calibration and functional testing were performed.The equipment was operational.
|