| Catalog Number |
426-2522X |
| Medical Device Problem Codes |
Break (1069); Stretched (1601); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, the pta balloon catheter was allegedly detached from the shaft when package was opened.The procedure was completed using another device.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records was performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: d2b (dqy; lit) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, after removing the balloon sleeve the balloon was allegedly detached from the catheter shaft.It was further reported that a break was confirmed at the balloon taper point and the catheter outer shaft.It was further reported that there was evidence of inner stretched within the break area.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the second complaint reported for this product/lot number combination.However, the device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.Images and video files were not provided for review.A balloon detachment was not present.However a break was confirmed at the balloon taper point and the catheter outer shaft.There was also evidence of inner stretched within the break area.The sleeve was returned present on the balloon and was removed.The inner diameter of the sleeve was measured and complied to specifications.Therefore, the result of the investigation is unconfirmed for the reported balloon detachment issue.The sleek otw device was returned for evaluation.The root cause for the reported balloon detachment issue could not be determined based upon the available information received from the field communications and device evaluation.Labeling review: the instructions for use for this sleek otw device (in122, rev.06) was reviewed and the following sections are applicable.Balloon characteristics: please check the package label for the rated burst pressure (rbp).It is important that the balloon not be inflated beyond the rated burst pressure.Pressures in excess of rated burst pressure may cause the balloon to burst.Compliance charts are provided with each product.Please note that balloon diameters may vary within manufacturing tolerances.All inflations should be viewed under fluoroscopy.The sleek® otw balloons reach their nominal diameter at 6 atm (608 kpa).Indications: the sleek® otw catheters are intended for balloon dilatation of the femoral, popliteal and infra-popliteal arteries.These catheters are not designed to be used in the coronary arteries.Warnings: contents supplied sterile using ethylene oxide (eo).Non-pyrogenic.Do not reuse, reprocess or resterilize.This product is designed and intended for single use.It is not designed to undergo reprocessing and re-sterilization after initial use.Reuse of this product, including after reprocessing and/or re sterilization, may cause a loss of structural integrity which could lead to a failure of the device to perform as intended and may lead to a loss of critical labeling/use information all of which present a potential risk to patient safety.Reusing this medical device bears the risk of cross-patient contamination as medical devices ¿ particularly those with long and small lumina, joints, and/or crevices between components ¿ are difficult or impossible to clean once body fluids or tissues with potential pyrogenic or microbial contamination have had contact with the medical device for an indeterminable period of time.The residue of biological material can promote the contamination of the device with pyrogens or microorganisms which may lead to infectious complications.Visually inspect the packaging to verify that the sterile barrier is intact.Do not use if the sterile barrier is opened or damaged.Use the catheter prior to the ¿use by¿ date specified on the package.Precautions: carefully inspect the catheter prior to use to verify that the catheter has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.Storage: store in a cool, dark, dry place.Use the catheter prior to the ¿use by¿ date specified on the package.Directions for use: inspection and preparation: remove the balloon sleeve by first withdrawing the shipping mandrel slightly and then slowly removing the sleeve while holding the catheter as close to the balloon as possible.If any resistance is felt, or if any stretching of the catheter is observed while removing the balloon sleeve, the product should not be used.The catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Prepare a mixture of contrast medium and normal saline as per normal procedure.(recommended 25%/75%).H10: h6 (device).H11: d2b: (dqy; lit), e1, h6 (method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the second complaint reported for this product/lot number combination.However, the device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sleek otw device was returned for evaluation.A balloon detachment was not present.However a break was confirmed at the balloon taper point and the catheter outer shaft.There was also evidence of inner stretched within the break area.The sleeve was returned present on the balloon and was removed.The inner diameter of the sleeve was measured and complied to specifications.There was no other damage noted to the balloon or the device.Therefore, the result of the investigation is unconfirmed for the reported balloon detachment issue.The root cause for the reported balloon detachment issue could not be determined based upon the available information received from the field communications and device evaluation.Labeling review: the instructions for use for this sleek otw device was reviewed and the following sections are applicable balloon characteristics: please check the package label for the rated burst pressure (rbp).It is important that the balloon not be inflated beyond the rated burst pressure.Pressures in excess of rated burst pressure may cause the balloon to burst.Compliance charts are provided with each product.Please note that balloon diameters may vary within manufacturing tolerances.All inflations should be viewed under fluoroscopy.The sleek otw balloons reach their nominal diameter at 6 atm (608 kpa).Indications: the sleek otw catheters are intended for balloon dilatation of the femoral, popliteal and infra popliteal arteries.These catheters are not designed to be used in the coronary arteries.Warnings: contents supplied sterile using ethylene oxide (eo).Non-pyrogenic.Do not reuse, reprocess or resterilize.This product is designed and intended for single use.It is not designed to undergo reprocessing and re sterilization after initial use.Reuse of this product, including after reprocessing and/or re sterilization, may cause a loss of structural integrity which could lead to a failure of the device to perform as intended and may lead to a loss of critical labeling/use information all of which present a potential risk to patient safety.Reusing this medical device bears the risk of cross patient contamination as medical devices ¿ particularly those with long and small lumina, joints, and/or crevices between components ¿ are difficult or impossible to clean once body fluids or tissues with potential pyrogenic or microbial contamination have had contact with the medical device for an indeterminable period of time.The residue of biological material can promote the contamination of the device with pyrogens or microorganisms which may lead to infectious complications.Visually inspect the packaging to verify that the sterile barrier is intact.Do not use if the sterile barrier is opened or damaged.Use the catheter prior to the use by date specified on the package.Precautions: carefully inspect the catheter prior to use to verify that the catheter has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.Storage: store in a cool, dark, dry place.Use the catheter prior to the use by date specified on the package.Directions for use: inspection and preparation: remove the balloon sleeve by first withdrawing the shipping mandrel slightly and then slowly removing the sleeve while holding the catheter as close to the balloon as possible.If any resistance is felt, or if any stretching of the catheter is observed while removing the balloon sleeve, the product should not be used.The catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Prepare a mixture of contrast medium and normal saline as per normal procedure.(recommended 25%/75%).H11: b5, d2b (dqy; lit).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, the pta balloon catheter was allegedly detached from the shaft when package was opened.The procedure was completed using another device.There was no patient contact.
|
| |
|
Search Alerts/Recalls
|
|