• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 6068073
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis # (b)(4):part # 6068073: lot # 0934823w visual and functional inspection confirmed the inner screw is jammed not allowing the implant to expand or collapse.It appears the screw has been over torqued causing the implant to jam.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider, distributor) regarding a patient having posterior lumbar interbody fusion l4/5 for lumbar canal stenosis.It was reported that the driver broke when an attempt was made to expand the cage.The tip of the inserter driver broke when the cage was expanded.One side of the cage in the facility was expanded to the left and right at the same time.As such, the cage was removed.The cage and the driver were replaced with a new one and installed.There was a possibility of screwdriver deterioration and a possibility that the cage has a defect (because the removed cage moved around/was loose).There was a delay in the procedure by less than 60 minutes.There was no patient symptom reported.There were no further complications reported regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21444839
Report Number1030489-2025-00965
Device Sequence Number17941221
Product Code MAX
UDI-Device Identifier00763000246693
UDI-Public00763000246693
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K210425
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6068073
Device Catalogue Number6068073
Device Lot Number0934823W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2023
Initial Report FDA Received Date02/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-