| Model Number |
714150 |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the button leaked after just 8 days of use.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Prior to a lot¿s release, the lot must be deemed acceptable by-passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.There were no samples returned for evaluation; therefore, the affected device could not be evaluated to confirm the reported failure mode.As such, a root cause could not be determined.However, the reported affected component is produced by an external supplier; our assembly process consists only of a pick and place operation for this material.A supplier corrective action request has been sent to the supplier to further investigate and address the reported condition.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation.The device was inspected, and the reported condition was observed.A supplier corrective action request has previously been sent to the supplier to further investigate and address the reported condition.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|
Search Alerts/Recalls
|