| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A non health care professional reported that during an intraocular lens (iol) implant procedure, reported experiencing several issues with the leading haptic stuck under the lens when being injected with company injector.Stated no negative impact to the patient but it required several extra minutes of operation room time to unstick the haptic.All processes were followed correctly in the loading process and it was advanced immediately prior to injection.Injectors are not available to return.Additional information has been requested.There are three medical device reports associated with this report.This file is 3 of 3.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.It was reported that haptic was observed to be stuck to the optic after implantation.The root cause of the reported issue is deemed to be related to manufacturing process change that increased the likelihood of company iol haptics adhering to the posterior optic surface following delivery with the company device.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|