| Model Number |
DDMB1D1 |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Shock from Patient Lead(s) (3162); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited undersensing during atrial tachycardia/atrial fibrillation (at/af) and the right ventricular (rv) lead displayed rare oversensing on stored electrograms (egm).In addition, the implantable cardioverter defibrillator (icd) delivered shocks for atrial fibrillation (af) with rapid ventricular response that was detected in the ventricular fibrillation (vf) zone.The device and leads remain in use.No further patient complications have been reported as a result of this event.
|
| |
|