• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO MULTIFIRE VERSATACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO MULTIFIRE VERSATACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174023
Medical Device Problem Codes Product Quality Problem (1506); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, in an open hernia repair, no tacks were able to be fired normally from the device.There was a noticeable gap between the two parts of the tip of the tacker.Another device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the e-piece was disengaged from the device.There was evidence of weld on the device.The sulu was received attached to the device with six tack remaining.It was reported that the instrument did not fire and there was product appearance.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur by not completing the firing cycle, and trying to remove the reload with excessive force resulting in the disengaged e-piece.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE VERSATACK
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21447463
Report Number2647580-2025-00612
Device Sequence Number7686770
Product Code GCJ
UDI-Device Identifier10884521070578
UDI-Public10884521070578
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K912097
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number174023
Device Catalogue Number174023
Device Lot NumberP3B0411
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2025
Supplement Date Received by Manufacturer05/01/2025
Initial Report FDA Received Date02/24/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-