| Catalog Number |
TCR75 |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent 2/24/2025 d4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an urological procedure surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 4/1/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 4/9/2025 d4 batch #101d05 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that the tcr75 reload was received along with its opened packaging and with no apparent damage.The reload had the proximal 6 drivers up without staples and the remaining drivers down with staples present.Also one partially formed staple was found in the pocket of the reload.The returned reload had the swing tab in the locked position.The cartridge reloads swing tab was reset in order to fire the cartridge.The cartridge was tested for functionality with a test device and fired without any difficulties.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It should be noted that the cartridge reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.In addition, failure to complete the stroke may result in an incomplete staple line.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 101d05, and no non-conformances were identified.
|
| |
|