Surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
|
H11.Additional narrative: updated d4, h4, and h6.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Svd [structural valve deterioration] is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|