Catalog Number CS-24301-E
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "the doctor found swg kinked during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
 
Additional Manufacturer Narrative
(b)(4) the customer returned one, opened cvc set including the guide wire within its advancer, an 18ga introducer needle, and an arrow raulerson syringe (ars) for analysis.Signs of use were observed on the returned components.The guide wire was observed to have two kinks towards the center of the body.The distal j-bend was slightly misshapen but intact.Microscopic examination confirmed the kinks in the guide wire body.Both welds were present and were observed to be full and spherical.The major kinks in the guide wire were located 297 mm and 318 mm from the distal tip.The overall length of the guide wire measured 602 mm, which is within the specification limits of 596-604 mm per guide wire product drawing.The outer diameter of the guide wire measured 0.798 mm, which is within the specification limits of 0.788-0.826 mm per guide wire product drawing.The guide wire was functionally tested per the product instructions for use (ifu).The ifu provided with this kit instructs the user, "advance guidewire into arrow raulerson syringe approximately 10 cm until it passes through syringe valves or into introducer needle." the guide wire was advanced through the returned 18ga introducer needle and ars to functionally test the guide wire.The undamaged portions of the guide wire passed through with little to no resistance.A manual tug test confirmed that both the distal and proximal welds were intact.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit describes suggested techniques to minimize the likelihood of guide wire damage during use.The instructions caution, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested." the report that the guide wire kinked during use was confirmed through complaint investigation of the returned sample.The guide wire had two kinks towards the center of the body.The returned guide wire met all relevant dimensional/functional requirements, and a device history record review did not identify any manufacturing related issues.Based on the condition of the guide wire and the report that the damage was observed during use, unintentional use error caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the doctor found swg kinked during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 16 GA X 8" (20 CM)
Common Device Name
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21456477
Report Number3006425876-2025-00192
Device Sequence Number7913518
Product Code FOZ
UDI-Device Identifier00801902102225
UDI-Public(01)00801902102225(11)240522(17)260417(10)71F24E0555(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K900263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/17/2026
Device Catalogue NumberCS-24301-E
Device Lot Number71F24E0555
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2025
Supplement Date Received by Manufacturer03/17/2025
Initial Report FDA Received Date02/25/2025
Supplement Report FDA Received Date03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-