MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM-ECLIPSE® PATIENT MODULE; STIMULATOR, ELECTRICAL, EVOKED RESPONSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM-ECLIPSE® PATIENT MODULE; STIMULATOR, ELECTRICAL, EVOKED RESPONSE Back to Search Results
Model Number OPM660
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa 624392.H3: product analysis found not able to confirm reported fault.Impedance of stim return channel was fluctuating.Electrode pcb failure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that pre-op, it was found that module was not delivering current.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® PATIENT MODULE
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key21456911
Report Number1045254-2025-00594
Device Sequence Number12672923
Product Code GWF
UDI-Device Identifier00613994870810
UDI-Public00613994870810
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K061639
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 02/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPM660
Device Catalogue NumberOPM660
Device Lot Number209844820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2022
Initial Report FDA Received Date02/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-