| Model Number |
610-2221010-01 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During disassembly of the device to determine the root cause, the technician observed damage consistent with mis-handling.The lcd display will be replaced to resolve the report.The device will be recertified and returned to the customer once the repair estimate has been approved.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device's display powered on white.Clinical data can not be seen.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|