As reported, approximately 2 weeks post the initial left reverse total shoulder arthroplasty, the patient was complaining of pain.On x-ray there was a calcar fracture, but was stable after opening up.The infection markers were up so a poly exchange and a washout was done with samples taken.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.
|
|
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.D10: a10012 - gps implant kit v2 03024224052 320-02-38 - rs expanded glenosphere 38mm, +4mm offset (b)(6), 321-52-10 - 3.2mm k-wire, thd, short 2 k-wires per pack (b)(6), 320-15-03 - rs glenoid plate post (b)(6), deg, left (b)(6), 531-55-88 - ergo gps 3.2mm drill kit sterile (b)(6), 320-15-05 - eq rev locking screw (b)(6), 320-10-00 - equinoxe reverse tray adapter plate tray +0 (b)(6), 320-20-00 - eq reverse torque defining screw kit (b)(6), 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm (b)(6), 531-78-20 - shouldr gps hex pins kit (b)(6), 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm (b)(6), 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm (b)(6), 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm (b)(6).
|
|
The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.A review of the sterile certificates and/or the final endotoxin test reports was performed.All lots were accepted to the requirements.The reason for the reported revision due to suspected infection and bone fracture cannot be conclusively determined; however, it may be related to an underlying patient condition and/or the surgical procedure.Correction: h6 component code.
|