| Model Number |
EPIQ DIAGNOSTIC ULTRASOUND SYSTEM |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the swivel mechanism of their epiq cvx ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.A philips service engineer reseated and secured the bearing to repair the system.A thorough investigation was performed to identify the root cause of the reported issue.The engineering team determined the failure was caused by a hardware issue in the articulating arm latching mechanism preventing the locking solenoid from fully engaging.This action was reported to fda per 21 cfr part 806 on 7/16/21.Reference corrections and removal report number 3019216-07/16/21-002-c.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough investigation was performed to determine the cause of the reported problem.It was determined that the application of adhesive was inadequate allowing for the bearing to become unsecured and preventing the locking solenoid from fully engaging.
|
| |
|
Event or Problem Description
|
|
It was reported the swivel mechanism of their epiq cvx ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.A philips service engineer reseated and secured the bearing to repair the system.A thorough investigation was performed to identify the root cause of the reported issue.The engineering team determined the failure was caused by a hardware issue in the articulating arm latching mechanism preventing the locking solenoid from fully engaging.This action was reported to fda per 21 cfr part 806 on 7/16/21.Reference corrections and removal report number 3019216-07/16/21-002-c.
|
| |
|
Search Alerts/Recalls
|