• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Medical Device Problem Codes Computer Software Problem (1112); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This regulatory report is being submitted due to retrospective review through capa 624392.Product analysis: performance data collected from the mobile programmer was received and analyzed.Analysis of the service log found the customer comment that the mobile programmer application crashed was not confirmed.Continuation of d10: product id 24967 lot# serial# (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the mobile programmer displayed a message indicating that the software had been updated and then the mobile pro grammer crashed.The user then interrogated a device during a demonstration and the mobile programmer indicated the device software had been updated and then an unexpected error was displayed.Troubleshooting steps were taken which involved uninstalling an reins talling the mobile programmer application however the issue persisted.Further troubleshooting advice was provided.The mobile programmer remains in use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK SMARTSYNC BASE
Common Device Name
ANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key21458418
Report Number2182208-2025-00987
Device Sequence Number20753172
Product Code DTC
UDI-Device Identifier00643169957657
UDI-Public00643169957657
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24970A
Device Catalogue Number24970A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Initial Report FDA Received Date02/25/2025
Date Device Manufactured12/07/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11...."
Patient SexUnknown
-
-