WILSON-COOK MEDICAL INC (COOK ENDOSCOPY) FUSION QUATTRO EXTRACTION BALLOON; CATHETER, BILIARY, SURGICAL
|
Back to Search Results |
|
| Model Number |
G31922 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E initial reporter phone: (b)(6).Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with the syringe still attached to the inflation port.The balloon was ruptured and socked over the distal tip.Under magnification, it was identified there was a portion where the balloon material did not match up and a portion appears to be missing.The missing portion of the balloon was not returned with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the report.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.In the information provided in the report, it states the balloon was tested prior to advancement down the endoscope accessory channel and inflated properly.A pinhole, split, or rupture in the balloon can occur if the balloon material has come into contact with a sharp object, such as a sharp stone or possibly a burr in the endoscope channel.A split or rupture in the balloon material can also occur if added pressure was applied during extraction.The instructions for use direct the user to "gently withdraw the inflated balloon toward the papilla." the instructions for use contain the following: ¿warning: do not exert excessive pressure on ampulla while extracting stones.If stone does not pass easily, reassess need for sphincterotomy.¿ prior to distribution, all fusion quattro extraction balloon are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography, the physician used a fusion quattro extraction balloon.It was reported that the balloon had burst inside the bile duct, so when doctor removed it from the endoscope and looked at it, there was no balloon material at all at the tip.When a replacement balloon was used to remove the stones, no fragments of the ruptured balloon material from the defective product were found, so it is possible that some may remain in the bile duct.However, because a sphincterotomy (est) had been performed and a wide opening had been secured, the doctor believes that any fragments of the ruptured balloon material will flow out and does not consider this a problem.As fragments of the ruptured balloon material would not show up on a ct scan, no scans have been taken.The detached balloon material remained inside the patient¿s body to pass naturally.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|
|
|